Precision notation for voltmeters The 2019 Stack Overflow Developer Survey Results Are In Announcing the arrival of Valued Associate #679: Cesar Manara Planned maintenance scheduled April 17/18, 2019 at 00:00UTC (8:00pm US/Eastern)What does it mean when multimeter accuracy is marked as: ±0,03%+10Digit?What does a half-digit means in case of accuracy?What does it mean when multimeter accuracy is marked as: ±0,03%+10Digit?Implications of INL on the accuracy and resolution of an ADCprecision and accuracy of a measurement systemAccuracy of various elements in circuitDC motor RPM accuracy/precisionHow to calculate 50Hz active power measurement accuracy from datasheet?Accuracy of a multimeter over 10 years periodConfusion about the meaning a parameterMultimeter temperature coefficient - Specified accuracyProblem understanding impedance measurement device accuracy format
Homework question about an engine pulling a train
Why don't hard Brexiteers insist on a hard border to prevent illegal immigration after Brexit?
Example of compact Riemannian manifold with only one geodesic.
How to make Illustrator type tool selection automatically adapt with text length
Identify 80s or 90s comics with ripped creatures (not dwarves)
Are spiders unable to hurt humans, especially very small spiders?
Single author papers against my advisor's will?
Was credit for the black hole image misappropriated?
Drawing vertical/oblique lines in Metrical tree (tikz-qtree, tipa)
Is it ethical to upload a automatically generated paper to a non peer-reviewed site as part of a larger research?
Loose spokes after only a few rides
What happens to a Warlock's expended Spell Slots when they gain a Level?
What do I do when my TA workload is more than expected?
How many cones with angle theta can I pack into the unit sphere?
Is an up-to-date browser secure on an out-of-date OS?
Mortgage adviser recommends a longer term than necessary combined with overpayments
60's-70's movie: home appliances revolting against the owners
Didn't get enough time to take a Coding Test - what to do now?
Make it rain characters
For what reasons would an animal species NOT cross a *horizontal* land bridge?
Why not take a picture of a closer black hole?
Python - Fishing Simulator
Working through the single responsibility principle (SRP) in Python when calls are expensive
Windows 10: How to Lock (not sleep) laptop on lid close?
Precision notation for voltmeters
The 2019 Stack Overflow Developer Survey Results Are In
Announcing the arrival of Valued Associate #679: Cesar Manara
Planned maintenance scheduled April 17/18, 2019 at 00:00UTC (8:00pm US/Eastern)What does it mean when multimeter accuracy is marked as: ±0,03%+10Digit?What does a half-digit means in case of accuracy?What does it mean when multimeter accuracy is marked as: ±0,03%+10Digit?Implications of INL on the accuracy and resolution of an ADCprecision and accuracy of a measurement systemAccuracy of various elements in circuitDC motor RPM accuracy/precisionHow to calculate 50Hz active power measurement accuracy from datasheet?Accuracy of a multimeter over 10 years periodConfusion about the meaning a parameterMultimeter temperature coefficient - Specified accuracyProblem understanding impedance measurement device accuracy format
.everyoneloves__top-leaderboard:empty,.everyoneloves__mid-leaderboard:empty,.everyoneloves__bot-mid-leaderboard:empty margin-bottom:0;
$begingroup$
From the technical details of a multimeter:
AC Voltage: - Range and Accuracy: 200mV ± (1.0%+2); 2V ± (0.7%+3); 20 ± (1.0%+3); 200/600V ± (1.2%+3)
What does the notation "± (0.7%+3)" mean? Specifically, what does "+3" mean?
accuracy voltmeter
$endgroup$
add a comment |
$begingroup$
From the technical details of a multimeter:
AC Voltage: - Range and Accuracy: 200mV ± (1.0%+2); 2V ± (0.7%+3); 20 ± (1.0%+3); 200/600V ± (1.2%+3)
What does the notation "± (0.7%+3)" mean? Specifically, what does "+3" mean?
accuracy voltmeter
$endgroup$
3
$begingroup$
You wrote it so where did the information come from?
$endgroup$
– Andy aka
Mar 8 at 10:17
$begingroup$
See this similar question and its answers.
$endgroup$
– Uwe
Mar 8 at 11:50
add a comment |
$begingroup$
From the technical details of a multimeter:
AC Voltage: - Range and Accuracy: 200mV ± (1.0%+2); 2V ± (0.7%+3); 20 ± (1.0%+3); 200/600V ± (1.2%+3)
What does the notation "± (0.7%+3)" mean? Specifically, what does "+3" mean?
accuracy voltmeter
$endgroup$
From the technical details of a multimeter:
AC Voltage: - Range and Accuracy: 200mV ± (1.0%+2); 2V ± (0.7%+3); 20 ± (1.0%+3); 200/600V ± (1.2%+3)
What does the notation "± (0.7%+3)" mean? Specifically, what does "+3" mean?
accuracy voltmeter
accuracy voltmeter
edited Mar 8 at 11:39
SamGibson
11.7k41739
11.7k41739
asked Mar 8 at 10:07
user31264user31264
1311
1311
3
$begingroup$
You wrote it so where did the information come from?
$endgroup$
– Andy aka
Mar 8 at 10:17
$begingroup$
See this similar question and its answers.
$endgroup$
– Uwe
Mar 8 at 11:50
add a comment |
3
$begingroup$
You wrote it so where did the information come from?
$endgroup$
– Andy aka
Mar 8 at 10:17
$begingroup$
See this similar question and its answers.
$endgroup$
– Uwe
Mar 8 at 11:50
3
3
$begingroup$
You wrote it so where did the information come from?
$endgroup$
– Andy aka
Mar 8 at 10:17
$begingroup$
You wrote it so where did the information come from?
$endgroup$
– Andy aka
Mar 8 at 10:17
$begingroup$
See this similar question and its answers.
$endgroup$
– Uwe
Mar 8 at 11:50
$begingroup$
See this similar question and its answers.
$endgroup$
– Uwe
Mar 8 at 11:50
add a comment |
2 Answers
2
active
oldest
votes
$begingroup$
What does the notation "± (0.7%+3)" mean? Specifically, what does "+3" mean?
The "+3" in that context, is the number of least-significant digits (LSD) on that range, which can be in error. This notation is one of a few "industry standard" ways of specifying measurement accuracy.
The accuracy specification you quoted is from the 2V range of whichever meter is in the question. Therefore using that as an example, the accuracy on a reading on the 2V range is between:
(reading +0.7% +3 LSD) and (reading -0.7% -3 LSD)
I found the Etekcity MSR-A600 meter has the same AC voltage specification as you quoted, so I'll use that one to illustrate LSD a little more. That is a 3.5 digit meter i.e. maximum count = 1999. The maximum count value is important for this next calculation.
For a 3.5 digit meter on the 2V range (actual maximum value shown is 1.999V), the LSD value is 1mV (0.001V).
So the accuracy calculation on that 3.5 digit meter's 2V range, is:
reading ±0.7% ±3mV
(where the 3mV is the 3 LSD, on that 2V range, on a 3.5 digit meter)
The value of the LSD will vary on other ranges. For example, on a 3.5 digit meter's 200V range (maximum value shown of 199.9V) the LSD value on that range is 0.1V.
The "percentage of reading" value is the gain accuracy; the "number of LSD" value is the offset accuracy. They relate to different parts of the measurement process and so have different units.
Some more explanations, for future reference and showing more "worked examples" for different meters:
- BK Precision: How to Read Accuracy Specifications
Tektronix (Keithley): Specs: How Can I Apply an Accuracy Specification on a Data Sheet to My Specific Measurement (PDF download)- EE.SE: What does it mean when multimeter accuracy is marked as: ±0,03%+10Digit?
$endgroup$
add a comment |
$begingroup$
A bit supplemental to the question, but that spec is incomplete. As the voltage being measured is AC (alternating current), bandwidth limitations apply. As they do to any circuit or oscilloscope.
I have a shiny new PROSTER VC99 meter. AC accuracy is stated as ±(0.8% + 5). It's cheap. But, that is only officially for frequencies of 40 - 400Hz. Whilst you think of your meter as for measuring the mains voltage (50-60Hz), that's fine. It's alluring though to think of the meter as a kinda substitute for an oscilloscope and use it to measure all sorts of AC voltages. That's wrong, as the metered AC voltage will drop into the higher frequencies. For reference, the VC99 response stays surprisingly flat till 4kHz (tested with function generator and oscilloscope). At 40kHz, it's quickly rolled off to only 15% of the true peak to peak.
So the moral of this answer is: stick it into your wall sockets by all means, but make sure to read the full spec if you're going to stick it into your HI-FI or radar installation.
$endgroup$
$begingroup$
Good point. The meter I used as an example in my answer (which I found using a search for the same specification given in the question) doesn't list the frequency range for which the specification applies, in its HTML webpage. However its full manual lists similar frequency limitations to your example in its accuracy specification: "40Hz ~ 400Hz, sine wave RMS (average response)."
$endgroup$
– SamGibson
Mar 8 at 15:34
add a comment |
Your Answer
StackExchange.ifUsing("editor", function ()
return StackExchange.using("schematics", function ()
StackExchange.schematics.init();
);
, "cicuitlab");
StackExchange.ready(function()
var channelOptions =
tags: "".split(" "),
id: "135"
;
initTagRenderer("".split(" "), "".split(" "), channelOptions);
StackExchange.using("externalEditor", function()
// Have to fire editor after snippets, if snippets enabled
if (StackExchange.settings.snippets.snippetsEnabled)
StackExchange.using("snippets", function()
createEditor();
);
else
createEditor();
);
function createEditor()
StackExchange.prepareEditor(
heartbeatType: 'answer',
autoActivateHeartbeat: false,
convertImagesToLinks: false,
noModals: true,
showLowRepImageUploadWarning: true,
reputationToPostImages: null,
bindNavPrevention: true,
postfix: "",
imageUploader:
brandingHtml: "Powered by u003ca class="icon-imgur-white" href="https://imgur.com/"u003eu003c/au003e",
contentPolicyHtml: "User contributions licensed under u003ca href="https://creativecommons.org/licenses/by-sa/3.0/"u003ecc by-sa 3.0 with attribution requiredu003c/au003e u003ca href="https://stackoverflow.com/legal/content-policy"u003e(content policy)u003c/au003e",
allowUrls: true
,
onDemand: true,
discardSelector: ".discard-answer"
,immediatelyShowMarkdownHelp:true
);
);
Sign up or log in
StackExchange.ready(function ()
StackExchange.helpers.onClickDraftSave('#login-link');
);
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Post as a guest
Required, but never shown
StackExchange.ready(
function ()
StackExchange.openid.initPostLogin('.new-post-login', 'https%3a%2f%2felectronics.stackexchange.com%2fquestions%2f426231%2fprecision-notation-for-voltmeters%23new-answer', 'question_page');
);
Post as a guest
Required, but never shown
2 Answers
2
active
oldest
votes
2 Answers
2
active
oldest
votes
active
oldest
votes
active
oldest
votes
$begingroup$
What does the notation "± (0.7%+3)" mean? Specifically, what does "+3" mean?
The "+3" in that context, is the number of least-significant digits (LSD) on that range, which can be in error. This notation is one of a few "industry standard" ways of specifying measurement accuracy.
The accuracy specification you quoted is from the 2V range of whichever meter is in the question. Therefore using that as an example, the accuracy on a reading on the 2V range is between:
(reading +0.7% +3 LSD) and (reading -0.7% -3 LSD)
I found the Etekcity MSR-A600 meter has the same AC voltage specification as you quoted, so I'll use that one to illustrate LSD a little more. That is a 3.5 digit meter i.e. maximum count = 1999. The maximum count value is important for this next calculation.
For a 3.5 digit meter on the 2V range (actual maximum value shown is 1.999V), the LSD value is 1mV (0.001V).
So the accuracy calculation on that 3.5 digit meter's 2V range, is:
reading ±0.7% ±3mV
(where the 3mV is the 3 LSD, on that 2V range, on a 3.5 digit meter)
The value of the LSD will vary on other ranges. For example, on a 3.5 digit meter's 200V range (maximum value shown of 199.9V) the LSD value on that range is 0.1V.
The "percentage of reading" value is the gain accuracy; the "number of LSD" value is the offset accuracy. They relate to different parts of the measurement process and so have different units.
Some more explanations, for future reference and showing more "worked examples" for different meters:
- BK Precision: How to Read Accuracy Specifications
Tektronix (Keithley): Specs: How Can I Apply an Accuracy Specification on a Data Sheet to My Specific Measurement (PDF download)- EE.SE: What does it mean when multimeter accuracy is marked as: ±0,03%+10Digit?
$endgroup$
add a comment |
$begingroup$
What does the notation "± (0.7%+3)" mean? Specifically, what does "+3" mean?
The "+3" in that context, is the number of least-significant digits (LSD) on that range, which can be in error. This notation is one of a few "industry standard" ways of specifying measurement accuracy.
The accuracy specification you quoted is from the 2V range of whichever meter is in the question. Therefore using that as an example, the accuracy on a reading on the 2V range is between:
(reading +0.7% +3 LSD) and (reading -0.7% -3 LSD)
I found the Etekcity MSR-A600 meter has the same AC voltage specification as you quoted, so I'll use that one to illustrate LSD a little more. That is a 3.5 digit meter i.e. maximum count = 1999. The maximum count value is important for this next calculation.
For a 3.5 digit meter on the 2V range (actual maximum value shown is 1.999V), the LSD value is 1mV (0.001V).
So the accuracy calculation on that 3.5 digit meter's 2V range, is:
reading ±0.7% ±3mV
(where the 3mV is the 3 LSD, on that 2V range, on a 3.5 digit meter)
The value of the LSD will vary on other ranges. For example, on a 3.5 digit meter's 200V range (maximum value shown of 199.9V) the LSD value on that range is 0.1V.
The "percentage of reading" value is the gain accuracy; the "number of LSD" value is the offset accuracy. They relate to different parts of the measurement process and so have different units.
Some more explanations, for future reference and showing more "worked examples" for different meters:
- BK Precision: How to Read Accuracy Specifications
Tektronix (Keithley): Specs: How Can I Apply an Accuracy Specification on a Data Sheet to My Specific Measurement (PDF download)- EE.SE: What does it mean when multimeter accuracy is marked as: ±0,03%+10Digit?
$endgroup$
add a comment |
$begingroup$
What does the notation "± (0.7%+3)" mean? Specifically, what does "+3" mean?
The "+3" in that context, is the number of least-significant digits (LSD) on that range, which can be in error. This notation is one of a few "industry standard" ways of specifying measurement accuracy.
The accuracy specification you quoted is from the 2V range of whichever meter is in the question. Therefore using that as an example, the accuracy on a reading on the 2V range is between:
(reading +0.7% +3 LSD) and (reading -0.7% -3 LSD)
I found the Etekcity MSR-A600 meter has the same AC voltage specification as you quoted, so I'll use that one to illustrate LSD a little more. That is a 3.5 digit meter i.e. maximum count = 1999. The maximum count value is important for this next calculation.
For a 3.5 digit meter on the 2V range (actual maximum value shown is 1.999V), the LSD value is 1mV (0.001V).
So the accuracy calculation on that 3.5 digit meter's 2V range, is:
reading ±0.7% ±3mV
(where the 3mV is the 3 LSD, on that 2V range, on a 3.5 digit meter)
The value of the LSD will vary on other ranges. For example, on a 3.5 digit meter's 200V range (maximum value shown of 199.9V) the LSD value on that range is 0.1V.
The "percentage of reading" value is the gain accuracy; the "number of LSD" value is the offset accuracy. They relate to different parts of the measurement process and so have different units.
Some more explanations, for future reference and showing more "worked examples" for different meters:
- BK Precision: How to Read Accuracy Specifications
Tektronix (Keithley): Specs: How Can I Apply an Accuracy Specification on a Data Sheet to My Specific Measurement (PDF download)- EE.SE: What does it mean when multimeter accuracy is marked as: ±0,03%+10Digit?
$endgroup$
What does the notation "± (0.7%+3)" mean? Specifically, what does "+3" mean?
The "+3" in that context, is the number of least-significant digits (LSD) on that range, which can be in error. This notation is one of a few "industry standard" ways of specifying measurement accuracy.
The accuracy specification you quoted is from the 2V range of whichever meter is in the question. Therefore using that as an example, the accuracy on a reading on the 2V range is between:
(reading +0.7% +3 LSD) and (reading -0.7% -3 LSD)
I found the Etekcity MSR-A600 meter has the same AC voltage specification as you quoted, so I'll use that one to illustrate LSD a little more. That is a 3.5 digit meter i.e. maximum count = 1999. The maximum count value is important for this next calculation.
For a 3.5 digit meter on the 2V range (actual maximum value shown is 1.999V), the LSD value is 1mV (0.001V).
So the accuracy calculation on that 3.5 digit meter's 2V range, is:
reading ±0.7% ±3mV
(where the 3mV is the 3 LSD, on that 2V range, on a 3.5 digit meter)
The value of the LSD will vary on other ranges. For example, on a 3.5 digit meter's 200V range (maximum value shown of 199.9V) the LSD value on that range is 0.1V.
The "percentage of reading" value is the gain accuracy; the "number of LSD" value is the offset accuracy. They relate to different parts of the measurement process and so have different units.
Some more explanations, for future reference and showing more "worked examples" for different meters:
- BK Precision: How to Read Accuracy Specifications
Tektronix (Keithley): Specs: How Can I Apply an Accuracy Specification on a Data Sheet to My Specific Measurement (PDF download)- EE.SE: What does it mean when multimeter accuracy is marked as: ±0,03%+10Digit?
edited Mar 8 at 18:02
answered Mar 8 at 11:37
SamGibsonSamGibson
11.7k41739
11.7k41739
add a comment |
add a comment |
$begingroup$
A bit supplemental to the question, but that spec is incomplete. As the voltage being measured is AC (alternating current), bandwidth limitations apply. As they do to any circuit or oscilloscope.
I have a shiny new PROSTER VC99 meter. AC accuracy is stated as ±(0.8% + 5). It's cheap. But, that is only officially for frequencies of 40 - 400Hz. Whilst you think of your meter as for measuring the mains voltage (50-60Hz), that's fine. It's alluring though to think of the meter as a kinda substitute for an oscilloscope and use it to measure all sorts of AC voltages. That's wrong, as the metered AC voltage will drop into the higher frequencies. For reference, the VC99 response stays surprisingly flat till 4kHz (tested with function generator and oscilloscope). At 40kHz, it's quickly rolled off to only 15% of the true peak to peak.
So the moral of this answer is: stick it into your wall sockets by all means, but make sure to read the full spec if you're going to stick it into your HI-FI or radar installation.
$endgroup$
$begingroup$
Good point. The meter I used as an example in my answer (which I found using a search for the same specification given in the question) doesn't list the frequency range for which the specification applies, in its HTML webpage. However its full manual lists similar frequency limitations to your example in its accuracy specification: "40Hz ~ 400Hz, sine wave RMS (average response)."
$endgroup$
– SamGibson
Mar 8 at 15:34
add a comment |
$begingroup$
A bit supplemental to the question, but that spec is incomplete. As the voltage being measured is AC (alternating current), bandwidth limitations apply. As they do to any circuit or oscilloscope.
I have a shiny new PROSTER VC99 meter. AC accuracy is stated as ±(0.8% + 5). It's cheap. But, that is only officially for frequencies of 40 - 400Hz. Whilst you think of your meter as for measuring the mains voltage (50-60Hz), that's fine. It's alluring though to think of the meter as a kinda substitute for an oscilloscope and use it to measure all sorts of AC voltages. That's wrong, as the metered AC voltage will drop into the higher frequencies. For reference, the VC99 response stays surprisingly flat till 4kHz (tested with function generator and oscilloscope). At 40kHz, it's quickly rolled off to only 15% of the true peak to peak.
So the moral of this answer is: stick it into your wall sockets by all means, but make sure to read the full spec if you're going to stick it into your HI-FI or radar installation.
$endgroup$
$begingroup$
Good point. The meter I used as an example in my answer (which I found using a search for the same specification given in the question) doesn't list the frequency range for which the specification applies, in its HTML webpage. However its full manual lists similar frequency limitations to your example in its accuracy specification: "40Hz ~ 400Hz, sine wave RMS (average response)."
$endgroup$
– SamGibson
Mar 8 at 15:34
add a comment |
$begingroup$
A bit supplemental to the question, but that spec is incomplete. As the voltage being measured is AC (alternating current), bandwidth limitations apply. As they do to any circuit or oscilloscope.
I have a shiny new PROSTER VC99 meter. AC accuracy is stated as ±(0.8% + 5). It's cheap. But, that is only officially for frequencies of 40 - 400Hz. Whilst you think of your meter as for measuring the mains voltage (50-60Hz), that's fine. It's alluring though to think of the meter as a kinda substitute for an oscilloscope and use it to measure all sorts of AC voltages. That's wrong, as the metered AC voltage will drop into the higher frequencies. For reference, the VC99 response stays surprisingly flat till 4kHz (tested with function generator and oscilloscope). At 40kHz, it's quickly rolled off to only 15% of the true peak to peak.
So the moral of this answer is: stick it into your wall sockets by all means, but make sure to read the full spec if you're going to stick it into your HI-FI or radar installation.
$endgroup$
A bit supplemental to the question, but that spec is incomplete. As the voltage being measured is AC (alternating current), bandwidth limitations apply. As they do to any circuit or oscilloscope.
I have a shiny new PROSTER VC99 meter. AC accuracy is stated as ±(0.8% + 5). It's cheap. But, that is only officially for frequencies of 40 - 400Hz. Whilst you think of your meter as for measuring the mains voltage (50-60Hz), that's fine. It's alluring though to think of the meter as a kinda substitute for an oscilloscope and use it to measure all sorts of AC voltages. That's wrong, as the metered AC voltage will drop into the higher frequencies. For reference, the VC99 response stays surprisingly flat till 4kHz (tested with function generator and oscilloscope). At 40kHz, it's quickly rolled off to only 15% of the true peak to peak.
So the moral of this answer is: stick it into your wall sockets by all means, but make sure to read the full spec if you're going to stick it into your HI-FI or radar installation.
answered Mar 8 at 13:25
Paul UszakPaul Uszak
3,56021843
3,56021843
$begingroup$
Good point. The meter I used as an example in my answer (which I found using a search for the same specification given in the question) doesn't list the frequency range for which the specification applies, in its HTML webpage. However its full manual lists similar frequency limitations to your example in its accuracy specification: "40Hz ~ 400Hz, sine wave RMS (average response)."
$endgroup$
– SamGibson
Mar 8 at 15:34
add a comment |
$begingroup$
Good point. The meter I used as an example in my answer (which I found using a search for the same specification given in the question) doesn't list the frequency range for which the specification applies, in its HTML webpage. However its full manual lists similar frequency limitations to your example in its accuracy specification: "40Hz ~ 400Hz, sine wave RMS (average response)."
$endgroup$
– SamGibson
Mar 8 at 15:34
$begingroup$
Good point. The meter I used as an example in my answer (which I found using a search for the same specification given in the question) doesn't list the frequency range for which the specification applies, in its HTML webpage. However its full manual lists similar frequency limitations to your example in its accuracy specification: "40Hz ~ 400Hz, sine wave RMS (average response)."
$endgroup$
– SamGibson
Mar 8 at 15:34
$begingroup$
Good point. The meter I used as an example in my answer (which I found using a search for the same specification given in the question) doesn't list the frequency range for which the specification applies, in its HTML webpage. However its full manual lists similar frequency limitations to your example in its accuracy specification: "40Hz ~ 400Hz, sine wave RMS (average response)."
$endgroup$
– SamGibson
Mar 8 at 15:34
add a comment |
Thanks for contributing an answer to Electrical Engineering Stack Exchange!
- Please be sure to answer the question. Provide details and share your research!
But avoid …
- Asking for help, clarification, or responding to other answers.
- Making statements based on opinion; back them up with references or personal experience.
Use MathJax to format equations. MathJax reference.
To learn more, see our tips on writing great answers.
Sign up or log in
StackExchange.ready(function ()
StackExchange.helpers.onClickDraftSave('#login-link');
);
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Post as a guest
Required, but never shown
StackExchange.ready(
function ()
StackExchange.openid.initPostLogin('.new-post-login', 'https%3a%2f%2felectronics.stackexchange.com%2fquestions%2f426231%2fprecision-notation-for-voltmeters%23new-answer', 'question_page');
);
Post as a guest
Required, but never shown
Sign up or log in
StackExchange.ready(function ()
StackExchange.helpers.onClickDraftSave('#login-link');
);
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Post as a guest
Required, but never shown
Sign up or log in
StackExchange.ready(function ()
StackExchange.helpers.onClickDraftSave('#login-link');
);
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Post as a guest
Required, but never shown
Sign up or log in
StackExchange.ready(function ()
StackExchange.helpers.onClickDraftSave('#login-link');
);
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Sign up using Google
Sign up using Facebook
Sign up using Email and Password
Post as a guest
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
Required, but never shown
3
$begingroup$
You wrote it so where did the information come from?
$endgroup$
– Andy aka
Mar 8 at 10:17
$begingroup$
See this similar question and its answers.
$endgroup$
– Uwe
Mar 8 at 11:50